A 510(k) must be submitted under 21 CFR 807.81 at least 90 days before the
device is first marketed. FDA answers with one of four actions: the device is
substantially equivalent to a predicate, it is not, FDA needs more information,
or no 510(k) is required (21 CFR 807.100(a)). Marketing may start only after
the SE order.
The order is not an approval. 21 CFR 807.97 states that an SE determination
does not in any way denote official approval of the device. What FDA reviews
is a comparison: its SE guidance of July 28, 2014, lays it out as a
flowchart with decision points, from the predicate through intended use and
technological characteristics to performance data.
| Traditional 510(k) | Special 510(k) | Abbreviated 510(k) |
|---|
| For what | any 510(k) that uses neither of the other two forms | a change to your own cleared device | a device for which FDA guidance, special controls or recognized standards exist |
| What carries it | full performance data for the comparison with the predicate | results from design controls, reviewable as a summary or risk analysis | summary reports or a declaration of conformity |
| What FDA says | decision goal: 90 FDA days (MDUFA V) | review generally within 30 days; unsuitable cases become a Traditional 510(k) | the same content under 21 CFR 807.87 as any 510(k) |
Since October 1, 2023, a 510(k) is submitted as an eSTAR. Instead of an
acceptance review under the refuse to accept guidance, FDA checks the eSTAR in
a technical screening, intended within 15 days. Substantive Interaction follows
and, where needed, an additional information request, during which the clock
stops.
Outside this page are the choice between 510(k), De Novo and PMA on the page
FDA Strategy & Pathways, the
question of whether a change requires a new 510(k), and preparation for FDA
inspections under the QMSR.